How Accurate is an Oscilloscope? - How to Read a Scope Datasheet 2/3

Опубликовано: 27 Январь 2025
на канале: Keysight Labs
2,171
56

why bandwidth isn't spec'd for digital channels
Click to subscribe! ► http://bit.ly/Scopes_Sub ◄

Full playlist:    • How to Read an Oscilloscope Data Sheet  

Questions/comments? Ask me on Twitter:
@DanielBogdanoff   / danielbogdanoff  

Learn more about using oscilloscopes:
http://oscilloscopelearningcenter.com

Check out the EEs Talk Tech podcast:
   • EEs Talk Tech - An Electrical Enginee...  

We start with oscilloscope DC vertical gain accuracy and the lack of an AC accuracy specification. Combining that with the scope's frequency response you can get a good picture of overall accuracy.

3:40
Oscilloscope channel-to-channel isolation - when you provide a stimulus on one channel, what happens on the other channels?

Offset range - how far up and down can you move signals?

4:00
Mixed signal oscilloscope's digital channel specifications

Bandwidth isn't specified for digital channels. Instead, the toggle rate is specified in seconds.

6:00
Horizontal specs

Timebase accuracy - is a warranted spec in ppm. As electronics age they start to become less accurate.

timebase delay range - how far can you horizontally offset a signal? If you're trying to collect more than 500s of data, you'll need to go with an external trigger in to trigger your oscilloscope.

delta time accuracy using cursors - a combination of lots of different accuracy factor.

Oscilloscope acquisition modes - Normal, High Res, Averaging, XY mode.

People don't use XY mode very often, but people who do are very passionate about it.